Measurement of the Ferrite Content in (Duplex) Steel and Weld Seams
Components found in industrial plants – whether chemical, energy, petrochemical or other – are often subject to heat, aggressive agents and high pressure. These conditions demand steel types that are extremely corrosion and acid resistant even at high temperatures. When austenitic steels are used, it is important to make sure the ferrite content of the weld seams is within strict norms, because only the optimal ferrite content can ensure the best corrosion protection. For this reason, some industries have set standards, specifications and regulations for ferrite content
During the welding of joints on e.g. boilers and pipelines made of austenitic steel, the heat input causes modifications in the crystal lattice structure which lead to the formation of ferrite. Weld seams that are poor in ferrite do not have as much yield strength, but too much ferrite reduces their fracture toughness, ductility and corrosion resistance, so it is important that the welding process produces just the right amount to ensure the optimum chemical and mechanical properties of the weldment.
With duplex steel in particular, the ferrite content in the heat affected zone can easily deviate from the target values, either due to unsuitable filler materials or through incorrect heat input or cooling during the welding. Only on-site spot measurements can provide assurance that the processing did not change the ferrite content at the expense of crucial mechanical or corrosion-resistance properties.
To meet these requirements FISCHER developed the Feritscope® FMP30 handheld instrument, which measures the ferrite content using the magnetic induction method and displays it either as percent ferrite content or as a WRC (Welding Research Council) ferrite number. The instrument can be outfitted with a variety of probes in special shapes such as axial, angled or for measuring inside centre holes.
All magnetisable structure sections are measured i.e., in addition to delta ferrite also strain-induced martensite, for example, or other ferritic phases.
It is suited for measurements according to the Basler-Standard and according to DIN EN ISO 17655. Areas of application are onsite measurements, e. g. of austenitic plating as well as weld seams in stainless steel pipes, containers, boilers or other products made of austenitic or duplex steel.
simple and quick measurement
It is easy to measure the ferrite content accurately when using the Feritscope®. Upon probe placement on the surface of the specimen, the reading is displayed automatically and stored in the instrument. The probe can also be placed onto hard to reach areas. For such applications, the instrument features an “external start” function to trigger the measurements with the push of a button. This is ideal for measurements in pipes, bore holes or grooves.
Finding weld seams in polished surfaces is made easy through the “continuous display” instrument function. When scanning the surface with the probe with this function enabled, the continuous readings are displayed only. A change in the ferrite content reading indicates that the weld seam has been found. For easy ferrite content measurements along a weld seam, the instrument offers the “continuous measurement capture” function. When scanning the weld seam with the probe positioned, the continuous readings are captured and stored. This provides a ferrite content profile along the weld seam.
Measurement influencing factors do not significantly affect the instrument. Ferrite content measurements can be carried out regardless of the substrate material properties starting at a plating thickness of 3 mm.
Corrective calibrations with customer-specific calibration standards or correction factors (included) can be used to take influences of the specimen shape (strong curvature), plating and substrate thicknesses into account. The calibration is always stored measurement-application specific in the respective application memory.
The instrument is designed to be robust and is well suited to on-site measurement as well as use in QA rooms and metallurgical laboratories. There is a large display and user-friendly menu system with USB and Bluetooth options. A large memory with sophisticated data handling and statistical evaluation of measurements is provided.
Calibration standards
To obtain comparable measurement results, the instruments must be adjusted or calibrated using standards that can be traced to internationally recognised secondary standards. For this purpose, the IIW (International Institute of Welding, UK) developed secondary standards that have been determined by the TWI (The Welding Institute, UK) according to the method described in DIN EN ISO 8249 and ANSI/AWS A4.2.
Helmut Fischer offers certified calibration standard sets that are traceable to the TWI secondary standards for corrective and master calibrations. The standards of the FISCHER calibration standard sets list in addition to the ferrite numbers FN also the % Fe values. Influences including the shape of the part to be measured (strong curvature, thickness of the ferrite-containing coating, etc.) can be considered through corrective calibrations with customer-specific calibration standards or through correction factors (included). The normalisation and corrective calibration are stored application-specific in the respective application memory of the instrument.
Magnetic Induction Method
The Feritscope® FMP30 measures according to the magnetic induction method. A magnetic field generated by a coil begins to interact with the magnetic portions of the specimen. The changes in the magnetic field induce a voltage proportional to the ferrite content in a second coil. This voltage is then evaluated. All magnetic portions in the otherwise non-magnetic structure are measured, i.e., in addition to delta ferrite and other ferritic portions also strain-induced martensite, for example.
A specific advantage of the magnetic induction method for measuring the ferrite content is that a sigma phase, i.e., a Fe-Cr precipitation, which has formed due to excess ferrite content and unfavourable cooling conditions, for example, is recognised correctly as a non-ferritic structural component. In comparison, erroneous interpretation of ferrite content is likely in a metallographic section where a sigma phase is not easily distinguished from a ferritic structure.
The instrument can be outfitted with a variety of probes in special shapes such as axial, angled or for measuring inside centre holes.
Thanks for sharing Valuable information about Feritscope FMP 30. If possible any further information/white paper using Feritscope FMP 30 instrument .
Best Regards,